The Compression Hip Screw with Trochanter Stabilizing Plate for Internal Fixation of Unstable Intertrochanteric Fractures
Jin Ho Cho
Journal of the Korean Fracture Society 2005;18(3):221-226.
DOI: https://doi.org/10.12671/jkfs.2005.18.3.221
Published online: June 16, 2016

Department of Orthopedic Surgery, Ilsan Paik Hospital, Inje University, Koyang, Korea.

osd07@ilsanpaik.ac.kr

Copyright © The Korean Fracture Society. All rights reserved

  • 26 Views
  • 0 Download
next
  • PURPOSE: To evaluate trochanter stabilizing plate and compression hip screw can prevent excessive impaction and cutting-out in unstable intertrochanteric fractures. MATERIALS AND METHODS: One hundred twenty-one patients with intertrochanteric fractures were treated between December 1999 and March 2002. Of the patients, twenty-four patients were treated with an additional TSP on the CHS. Nineteen patients were followed for more than six months. The fractures were classified according to the AO classification. Impactions of compression lag screw were measured immediate postoperatively and postoperatively three months later on simple radiographs. Functional results were followed Salvati-Wilson assessment score at postoperative six months later. RESULTS: The group consisted of seven men and twelve women, and the mean age was 73 years. Two were classified as A2.2, six A2.3, one A3.2, and ten A3.3 fractures. Mean impaction was 5.4 mm (range 1.8 to 11.4 mm). Functional results were excellent in 32% and good in 53%. Eighteen patients had healed after operation. One complication required a bipolar hemiarthroplasty due to cutting-out of lag screw. CONCLUSION: In three-part and four-part intertrochanteric fractures with lateral cortex breakage or vertical fracture in greater trochanter, the addition of TSP to CHS can prevent abductor muscle weakness due to fracture impaction, limb shortening and additional lateral cortical fracture. It also helps early weight bearing and bone healing.

Figure & Data

REFERENCES

    Citations

    Citations to this article as recorded by  

      • Cite
        CITE
        export Copy Download
        Close
        Download Citation
        Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

        Format:
        • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
        • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
        Include:
        • Citation for the content below
        The Compression Hip Screw with Trochanter Stabilizing Plate for Internal Fixation of Unstable Intertrochanteric Fractures
        J Korean Fract Soc. 2005;18(3):221-226.   Published online July 31, 2005
        Close
      • XML DownloadXML Download
      We recommend
      Related articles
      The Compression Hip Screw with Trochanter Stabilizing Plate for Internal Fixation of Unstable Intertrochanteric Fractures
      The Compression Hip Screw with Trochanter Stabilizing Plate for Internal Fixation of Unstable Intertrochanteric Fractures

      J Musculoskelet Trauma : Journal of Musculoskeletal Trauma
      Close layer
      TOP